Fast Detector of the Powder Diffraction Station

Mythen detector (PSI Detector group/Dectris)

6 Modules (Dectris/PSI Detectors group)

Strip-pitch 50µm

Radius  550mm

Covered Angular Range 40˚

Energy Range 8 - 30keV

 

MAD26: High Resolution Detector of the Powder Diffraction Station

13 Channels with Si 111 and Si 220 crystals

Crystals can be changed via translation

Channel separation 1.5˚

Energy range 8-50keV

Exchangeable slits

 

Design and construction of Multicrystal Analyser Detectors using Rowland circles: application to MAD26 at Alba
I. Peral, J. McKinlay, M. Knapp and S. Ferrer
J. Synchrotron radiation  (2011) 18 , 1-9

 

 

Area Detector of the High  Diffraction Station

CCD camera, SX165 (Rayonix)                                                           

Area: round, 165mm diameter             rayonix

Energy range:  20- 50keV

Dynamic range: 16 bit

PSF(Point Spread Function): 200 μm FWHM

Read Noise: 9 e-/pixel @ 3.5 sec. readout, 

                     13 e-/pixel @ 2.5 sec. readout

                                                                   Dark Current <0.01 e-/pixel/sec.      

 